Intermittent mechanical ventilation


Intermittent Mandatory Ventilation (IMV) refers to any mode of mechanical ventilation where a regular series of breaths are scheduled but the ventilator senses patient effort and reschedules mandatory breaths based on the calculated need of the patient. Similar to continuous mandatory ventilation in parameters set for the patients pressures and volumes but distinct in its ability to support a patient by either supporting their own effort or providing support when patient effort is not sensed. IMV is frequently paired with additional strategies to improve weaning from ventilator support or to improve cardiovascular stability in patients who may need full life support.

Synchronized Intermittent Mechanical Ventilation is a variation of IMV, in which the ventilator breaths are synchronized with patient inspiratory effort. SIMV, with and without pressure support has not been shown to have any advantages over continuous mandatory ventilation (CMV) in terms of mortality or weaning success, and has been shown to result in longer weaning times when compared to t-piece trials or gradual reductions in pressure support. Some studies have shown an increase in patient work of breathing when switched from CMV to SIMV, and others have demonstrated potential detrimental effects of SIMV on respiratory muscles and respiratory drive.

Mandatory minute ventilation is a mode which requires the operator to determine what the appropriate minute ventilation for the patient should be, and the ventilator then monitors the patient's ability to generate this volume every 7.5 seconds. If the calculation suggests the volume target will not be met, SIMV breaths are delivered at the targeted volume to achieve the desired minute ventilation. Allows spontaneous breathing with automatic adjustments of mandatory ventilation to the meet the patient’s preset minimum minute volume requirement. If the patient maintains the minute volume settings for VT x f, no mandatory breaths are delivered. If the patient's minute volume is insufficient, mandatory delivery of the preset tidal volume will occur until the minute volume is achieved. The method for monitoring whether or not the patient is meeting the required minute ventilation (VE) is different per ventilator brand and model, but generally there is a window of time being monitored and a smaller window being checked against that larger window (i.e., in the Dräger Evita® line of mechanical ventilators there is a moving 20-second window and every 7 seconds the current tidal volume and rate are measured against to make a decision for if a mechanical breath is needed to maintain the minute ventilation). MMV is the most optimal mode for weaning in neonatal and pediatric populations and has been shown to reduce long term complications related to mechanical ventilation.


...
Wikipedia

...