Gaseous ionization detectors


Gaseous ionization detectors are radiation detection instruments used in particle physics to detect the presence of ionising particles, and in radiation protection applications to measure ionizing radiation.

They use the ionising effect of radiation upon a gas-filled sensor. If a particle has enough energy to ionize a gas atom or molecule, the resulting electrons and ions cause a current flow which can be measured.

Gaseous ionisation detectors form an important group of instruments used for radiation detection and measurement. This article gives a quick overview of the principal types, and more detailed information can be found in the articles on each instrument. The accompanying plot shows the variation of ion pair generation with varying applied voltage for constant incident radiation. There are three main practical operating regions, one of which each type utilises.

The three basic types of gaseous ionization detectors are:

All of these have the same basic design of two electrodes separated by air or a special fill gas, but each uses a different method to measure the total number of ion-pairs that are collected. The strength of the electric field between the electrodes and the type and pressure of the fill gas determines the detector's response to ionizing radiation.

Ionization chambers operate at a low electric field strength, selected such that no gas multiplication takes place. The ion current is generated by the creation of "ion pairs", consisting of an ion and an electron. The ions drift to the cathode whilst free electrons drift to the anode under the influence of the electric field. This current is independent of the applied voltage if the device is being operated in the "ion chamber region". Ion chambers are preferred for high radiation dose rates because they have no "dead time"; a phenomenon which affects the accuracy of the Geiger Muller tube at high dose rates.

The advantages are:

The disadvantages are:

Proportional counters operate at a slightly higher voltage, selected such that discrete avalanches are generated. Each ion pair produces a single avalanche so that an output current pulse is generated which is proportional to the energy deposited by the radiation. This is in the "proportional counting" region. The term "gas proportional detector" (GPD) is generally used in radiometric practice, and the property of being able to detect particle energy is particularly useful when using large area flat arrays for alpha and beta particle detection and discrimination, such as in installed personnel monitoring equipment. The Wire chamber is a multi-electrode form of proportional counter used as a research tool.


...