Bioelectrical impedance analysis


Bioelectrical impedance analysis (BIA) is a commonly used method for estimating body composition, and in particular body fat. Since the advent of the first commercially available devices in the mid-1980s the method has become popular owing to its ease of use, portability of the equipment and its relatively low cost compared to some of the other methods of body composition analysis. It is familiar in the consumer market as a simple instrument for estimating body fat. BIA actually determines the electrical impedance, or opposition to the flow of an electric current through body tissues which can then be used to estimate total body water (TBW), which can be used to estimate fat-free body mass and, by difference with body weight, body fat.

Many of the early research studies showed that BIA was quite variable and it was not regarded by many as providing an accurate measure of body composition. In recent years technological improvements have made BIA a more reliable and therefore more acceptable way of measuring body composition. Nevertheless, it is not a "gold standard" or reference method.

Although the instruments are straightforward to use, careful attention to the method of use (as described by the manufacturer) should be given.

Simple devices to estimate body fat, often using BIA, are available to consumers as body fat meters. These instruments are generally regarded as being less accurate than those used clinically or in nutritional and medical practice. They tend to under-read body fat percentage.

Dehydration is a recognized factor affecting BIA measurements as it causes an increase in the body's electrical resistance, so has been measured to cause a 5 kg underestimation of fat-free mass i.e. an overestimation of body fat.

Body fat measurements are lower when measurements are taken shortly after consumption of a meal, causing a variation between highest and lowest readings of body fat percentage taken throughout the day of up to 4.2% of body fat.

Moderate exercise before BIA measurements lead to an overestimation of fat-free mass and an underestimation of body fat percentage due to reduced impedance. For example, moderate intensity exercise for 90–120 minutes before BIA measurements causes nearly a 12 kg overestimation of fat-free mass, i.e. body fat is significantly underestimated. Therefore, it is recommended not to perform BIA for several hours after moderate or high intensity exercise.


...