The ankle-brachial pressure index (ABPI) or ankle-brachial index (ABI) is the ratio of the blood pressure at the ankle to the blood pressure in the upper arm (brachium). Compared to the arm, lower blood pressure in the leg is an indication of blocked arteries due to peripheral artery disease (PAD). The ABPI is calculated by dividing the systolic blood pressure at the ankle by the systolic blood pressure in the arm.

The patient must be placed supine, without the head or any extremities dangling over the edge of the table. Measurement of ankle blood pressures in a seated position will grossly overestimate the ABI (by approximately 0.3).

A Doppler ultrasound blood flow detector, commonly called Doppler wand or Doppler probe, and a sphygmomanometer (blood pressure cuff) are usually needed. The blood pressure cuff is inflated proximal to the artery in question. Measured by the Doppler wand, the inflation continues until the pulse in the artery ceases. The blood pressure cuff is then slowly deflated. When the artery's pulse is re-detected through the Doppler probe the pressure in the cuff at that moment indicates the systolic pressure of that artery.

The higher systolic reading of the left and right arm brachial artery is generally used in the assessment. The pressures in each foot's posterior tibial artery and dorsalis pedis artery are measured with the higher of the two values used as the ABI for that leg.

The ABPI test is a popular tool for the non-invasive assessment of PVD. Studies have shown the sensitivity of ABPI is 90% with a corresponding 98% specificity for detecting hemodynamically significant (Serious) stenosis >50% in major leg arteries, defined by angiogram.


...