AutoAnalyzer


The AutoAnalyzer is an automated analyzer using a flow technique called continuous flow analysis (CFA), first made by the Technicon Corporation. The instrument was invented 1957 by Leonard Skeggs, PhD and commercialized by Jack Whitehead's Technicon Corporation. The first applications were for clinical analysis, but methods for industrial analysis soon followed. The design is based on separating a continuously flowing stream with air bubbles.

In continuous flow analysis (CFA) a continuous stream of material is divided by air bubbles into discrete segments in which chemical reactions occur. The continuous stream of liquid samples and reagents are combined and transported in tubing and mixing coils. The tubing passes the samples from one apparatus to the other with each apparatus performing different functions, such as distillation, dialysis, extraction, ion exchange, heating, incubation, and subsequent recording of a signal. An essential principle of the system is the introduction of air bubbles. The air bubbles segment each sample into discrete packets and act as a barrier between packets to prevent cross contamination as they travel down the length of the tubing. The air bubbles also assist mixing by creating turbulent flow (bolus flow), and provide operators with a quick and easy check of the flow characteristics of the liquid. Samples and standards are treated in an exactly identical manner as they travel the length of the tubing, eliminating the necessity of a steady state signal, however, since the presence of bubbles create an almost square wave profile, bringing the system to steady state does not significantly decrease throughput ( third generation CFA analyzers average 90 or more samples per hour) and is desirable in that steady state signals (chemical equilibrium) are more accurate and reproducible.

A continuous flow analyzer (CFA) consists of different modules including a sampler, pump, mixing coils, optional sample treatments (dialysis, distillation, heating, etc.), a detector, and data generator. Most continuous flow analyzers depend on color reactions using a flow through photometer, however, also methods have been developed that use ISE, flame photometry, ICAP, fluorometry, and so forth.

Flow injection analysis (FIA), was introduced in 1975 by Ruzicka and Hansen, The first generation of FIA technology, termed flow injection (FI), was inspired by the AutoAnalyzer technique invented by Skeggs in early 1950s. While Skeggs' AutoAnalyzer uses air segmentation to separate a flowing stream into numerous discrete segments to establish a long train of individual samples moving through a flow channel, FIA systems separate each sample from subsequent sample with a carrier reagent. While the AutoAnalyzer mixes sample homogeneously with reagents, in all FIA techniques sample and reagents are merged to form a concentration gradient that yields analysis results.


...
Wikipedia

...